Hot search: tipping  cigarette  dental 
 

Central Dialysis Reverse Osmosis Water Systems

Click on the image to check the artwork
unit price: negotiable
min amount:
amount:
period for delivery: Consignment Deadline Days
area: Sichuan
Expiry date : Long Effective
last update: 2021-09-15 06:19
view count: 349
enquiry
Company Profile
 
 
Product details

Our central dialysis reverse osmosis water systems purified water is in compliance with registered enterprise product standards, USA AAM/ASIO standard and YY0572-2015 Standard. System design strictly compliance with YY0793-2010 standard .

It can be used with dialysis machines such as Fresenius, Braun, Campbell, Baxter, Toray, Nikkiso, Nipro, Belk, and Jihua.


System Features

● Removal rate of endotoxin and bacteria ≥99%

● Dissolved salt removal rate ≥97%

● The water recovery rate is more than 70%

● System emptying rate ≥99%, no dead space


Models for central dialysis reverse osmosis systems

Model

Capacity

Feed for dialysis machine

Moldialysis50D

50LPH

1set

Moldialysis250D

250LPH

3-4sets

Moldialysis500D

500LPH

5-10sets

Moldialysis750D

750LPH

11-15sets

Moldialysis1000D

1000LPH

15-20sets

Advantages

1. Fully automatic control , failure indicator and onsite alarm.

2. Built-in automatic water circulation program to maintain water quality.

3. Built-in timing flushing program, RO membrane anti-scaling program, and RO membrane flushing function.

4. The touch screen has a process picture, and each water pump, switch, and liquid level has a dynamic monitoring picture.

5. The system has automatic protection functions, such as water stop, power failure, overload, high and low voltage automatic protection to ensure equipment safety.



Photos of 3000LPH dialysis reverse osmosis water systems


Photo of 500LPH dialysis reverse osmosis water system


FAQ

Fault content

Problems

Solutions



Produced water conductivity over standard

RO membrane fouling

Replace RO membrane

Raw water ion content high

Test the raw water and produced water conductivity


Produce water conductivity meter failure

Replace or maintance the produce water conductivity meter


Concentrated water discharge valve failure

Replace or maintance the concentrated water discharge valve




Low production water flow

RO membrane fouling

Replace ro membrane

Low Raw water temperature

Increase inlet water temperature,15-25℃


Low Running pressure

Increase running pressure, less than 200bar


Low ambient temperature

Increase the temp around 20℃




http://www.mole-water.com/

Total0bar [View All]  Related Comments
 
More>other products

[ ProductsSearch ]  [ Send Fav ]  [ Share ]  [ Print ]  [ Send Report ]  [ Close ]

 
Home | Products | Suppliers | News | Tradeshow | Sitemap | Message | RSS Feed